Analytical Microwave Fire
1891 Views
Feedback
1 Comments

Favorite
Image for Analytical Microwave Fire

Type:  Lessons Learned

Publisher:  Idaho National Laboratory, Idaho Falls, ID

Published As:  Public

Date: 

Topics:  Electrical, Fire Protection, Laboratory, Hazardous Energy Control

Lessons Learned Recommendations:
1. Ensure the temperature probe is compatible with the microwave model being used. Consult the manufacturer if there is any doubt.
2. Remove incompatible parts/consumables from the equipment operational area.
3. For accurate temperature measurements during a microwave run, the thermocouple needs to measure the temperature of the liquid phase during the run. The minimum liquid volume needed per 100 ml reaction vessel is stated as 5 ml acid by the manufacturer when using a compatible thermocouple in an EasyPrep sample vessel.
4. Consider use of the exhaust acid/organic vapor sensor for unattended microwave runs.