Failures of Lapel Samplers from SKC Model XR5000
2772 Views
Feedback
0 Comments

Favorite
Image for Failures of Lapel Samplers from SKC Model XR5000

Type:  Just-In-Time Report

Publisher:  CH2MHill Plateau Remediation Company (Richland, WA - Hanford Site), Richland, WA - Hanford Site (CHPRC)

Published As:  Public

Date: 

Topics:  Respiratory Protection, Radiation Protection

Several failures of SKC Model XR5000 lapel samplers have occurred during entries into contamination areas. The investigation identified two reasons for the failures. A manufacturing defect caused some lithium-ion batteries to fail. Also, during the maintenance check of newly received lapel samplers, instances were found where the screw connecting the battery to the unit was too loose or overly tight resulting in the battery separating from the unit while being used in the field.

You must be logged in to open the attachment.

Download Article
Comments about this Article
You must be logged in to add or view comments.
Related Articles
Feedback Form
How did you apply this information?
You must be logged in to use this feature
Is this article applicable to your organization?